Air tightness of box body | 0.1% chamber volume at + 1000 and-1000Pa |
Box resistance to pressure | ≥2500Pa |
filter efficiency | ≥99.995%@MPPS |
Filter pickup mode | In-situ manual / automatic scanning for leak collection; |
Deal with air volume | 240m³ / h-2000m³ / h (according to different models); |
Resistance detection method | Real-time monitoring of mechanical type differential pressure meter or variable type differential pressure sensor |
Disinfection method of the filter | Can be sterilized in situ |
resistance to corrosion | Made of high quality stainless steel, inside and outside full welding disinfectant, detergent and acid, alkali |
High efficiency particulate air filter | Air volume (0.9 m/s) | Box size | The wind flange | |
circular | rectangle | |||
HxWxD | CHM | HxWxD | ΦD | AxB |
305X305X93 | 300 | 735X465X450(+60) | 150 | 200X150 |
610X305X93 | 600 | 735X770X450(+60) | 200 | 200X200 |
610X530X93 | 1000 | 1040X690X450(+60) | 250 | 320X200 |
610X610X93 | 1250 | 1040X770X450(+60) | 300 | 320X250 |
610X762X93 | 1500 | 1040x922X450(+60) | 300 | 320X320 |
610X762X117 | 2000 | 1040X922X470(+60) | 350 | 400X320 |